功能

  • It can be used for explosion prevention control and constant monitoring of hydrogen gas in research and laboratory.
  • Hydrogen gas concentration is displayed by two methods, bar meter and digital.
  • The name of the hydrogen gas changes to three colors to indicate the detection status, so that the detection status can be visually confirmed from a distance (three colors: normal: green, 1st alarm: orange, 2nd alarm: red).
  • The indication alarm unit (model: GP-6001) adopts an independent unit for easy installation.
  • The two-stage alarm function enables the gradual management of gas alarms.
  • Adoption of maintenance mode function makes it possible to shut off external alarm during maintenance.
  • Gas detection part: Compact and lightweight, pressure-resistant explosion-proof structure (explosion-proof, etc., so it can be used in hydrogen atmosphere), dustproof and waterproof structure (protection grade: equivalent to IP 67), and can be installed in harsh environments.

规格

  • Detection principle: catalytic combustion type
  • Target gas: Hydrogen
  • Type of indicating alarm part: GP-6001
  • Indication alarm part structure: Wall mounted type
  • Explosion-proof Gas Detector: Explosion-proof Pressure
  • Type of gas detector: GD-A80
  • Display text: H2
  • Sensor Type: HW-6211
  • Detection range: 0 ~ 100%LEL
  • Alarm Setting: 1st Stage: 25%LEL/2nd Stage: 50%LEL
  • Number of detection points: 1 point
  • Alarm accuracy: within ±25% of the alarm setting value
  • Alarm delay time: Within 30 seconds when a gas 1.6 times the alarm setting value is detected.
  • Gas alarm type: H-HH alarm
  • gas alarm operation
  • Self-holding operation (automatic return after reset)
  • 1st Stage: ALM1 Lamp Flashing or Lit (Red), Buzzer
  • 2nd Stage: ALM2 Lamp Flashing or Lit (Red), Buzzer
  • Contact output (1st Stage, 2nd Stage Individual, No Voltage)
  • failure alarm operation
  • automatic return operation
  • FAULT Lamp Flashing (Orange), Contents Display, Buzzer
  • Contact output (no voltage)
  • Contact Specifications: Non-excitation at Normal Time/a Contact (Open at Normal Time)
  • Contact Capacity: 125 VAC/1 A, 30 VDC/1 A for Each Output (In the case of inductive load, install an appropriate spark killer on the load.)
  • Input power: AC type (100 VAC, 50/60Hz)

相关产品推荐

您可能还对以下产品感兴趣

1-5237-15 风速风速计516.0210箱 516.021
理化检测
61-6497-48 剪辑适配器 CL-14
理化检测

61-6497-48 剪辑适配器 CL-14

型号: 61-6497-48 品牌: SANWA ELECTRIC INSTRUMENT CO., LTD. (Sanwa)
64-0900-94 数字照度计 CHE-LT1
理化检测

64-0900-94 数字照度计 CHE-LT1

型号: 64-0900-94 品牌: SANWA SUPPLY INC.
1-7996-01 [已停用]小型氧气监视器集成 OX-500-00
理化检测