理化检测
67-4914-17 Mg1000Series Wall-Mounted Oxygen Meter (Transmission Signal: 0~250 Mv, Acidity Meter Mode Ii, Display: 0.0~25.0%O2 Fixed, Transmission Output: 0.0~25.0%O2 Supported) MG1000-100
产品描述
产品型号: 67-4914-17产品品牌: CHINO CORPORATION 功能 Oxygen detec […]
联系我们
产品详细内容
产品型号: 67-4914-17
产品品牌: CHINO CORPORATION
功能
- Oxygen detector and indicator are compactly assembled, and alarm output function and external output for connection such as recorder are equipped as standard.
- The MG1000 series is a wall-mounted oximeter that uses a limit-current zirconia oximeter. It is a digital oximeter that can measure oxygen concentration from a low concentration as an oxygen-deficiency meter or oximeter, with various alarm functions as well as display of measured values.
- High performance and small size.
- Uses long-life zirconia oxygen sensing element to maintain high precision and excellent reproducibility.
- Equipped with analog output and alarm relay output as standard.
- Two models are available: a built-in detector and a separate detector.
- zirconia oximeter
- Stabilized zirconia is known as an oxygen ion conductor.
- The stabilized zirconia is used as a substrate, and a platinum electrode and an oxygen diffusion limiting hole are formed on one surface of the substrate, and a platinum heater is formed on the other surface.
- In order to obtain practical oxygen ion conductivity, the substrate is heated to several hundred °C by energizing the heater. When a voltage is applied between the electrodes, a current with oxygen ions as carriers flows through the substrate.
- Due to the oxygen diffusion limiting hole formed on the cathode, the amount of oxygen flowing into the cathode is limited, and a flat region (limit current) is observed in the current-voltage characteristics.
- Since the limiting current varies in proportion to the oxygen concentration, it is possible to determine the oxygen concentration in the atmosphere by measuring the limiting current.
规格
- Detector Mounting Position: Built-in
- Separation detection part cable length: None
- Transmission signal: 0~250 mV
- alarm setting
- Anoxia Meter Mode Type II
- For anoximeter mode Type I
- Only the lower limit value can be specified within the range of 18~21%.
- Display range: 0.0~25.0%O2 fixed
- Transmission output range: output corresponding to 0.0~25.0%O2
- *The display range and transmission output range can be combined in the following formats: 00, 11, 20, 22. However, only 00 can be specified when the alarm setting is in the oxygen deficiency meter mode.
- In the case of display range 1 in the moat meter mode Type I, the initial alarm value is lower limit 0.00%, upper limit 2.50%.
- Measurement Principle: Zirconia Solid Electrolyte Method
- Air sampling method: natural diffusion type
- Response time: 90% response within 20 seconds
- Oxygen detection element position: fixed to the dedicated holder in the main body case
- Measurement value display: Digital display with 3-digit yellow-green LED
- Measuring range (resolution): 0.0~25.0 (0.1)%O2
- Accuracy rating: +/- 0.5%O2 +/- 1digit in a dry gas of oxygen and nitrogen
- Transmission method: output to 0.0~25.0 measurement indications
- Alarm function: Measured value alarm ...
- Alarm Setting Code: Anoxic Meter Mode
- Lower warning value: specified within 18~21%
- Upper warning value: fixed at 26%
- Relay power yes/no: yes
- Buzzer with/without: Yes
- Lamp lit/not lit: yes
- Output holding yes/no: yes
- Detection element abnormality alarm: When the heater current to the oxygen detection element is abnormal, relay output is ON, buzzer is blown, and lamp is lit regardless of the content of the measured value alarm.
- Power failure warning: When the power supply to the tester is OFF or when the fuse is broken, relay output is ON.
- Alarm relay output: number of output circuits (2), number of output circuits (1) in case of detectors separated type, output capacity (Resistive Load 110 V AC 0.2 A, 24 V DC 1 A/Inductive Load 110 V AC 0.1 A, 24 V DC 0.5 A)
- Warm-up time: 5 minutes
- Note)Warning and transmission signals are not output during warm-up.
- Operating Temperature Range: 0~40°C
- Operating humidity range: 80%rh or less
- Power: 100 VAC ±10% 50/60Hz
- Power Consumption: Approx. 8VA
- Weight: Body/Approx. 500 g
- Mounting method: wall or switchbox mounting
- ISO 14001 Certified JQA-EM2414
- ISO 9001 Certification JQA-0656
相关产品推荐